| SENSORS |
| SENSOR |
1000 |
2000 |
2500 |
3000 |
3500 |
3500
DT |
4000 |
4200 |
4250 |
4300 |
4400 |
4500 |
4500
HOR |
ACCURACY (+/-)* |
RESOLUTION |
SPECIFICATION RANGE |
OPERATIONAL RANGE |
NOTES |
| Wind Speed | Air Flow |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
Larger of 3% of reading, least significant digit or
20 ft/min |
0.1 m/s
1 ft/min
0.1 km/h
0.1 mph
0.1 knots
1 B |
0.6 to 40.0 m/s
118 to 7,874 ft/min
2.2 to 144.0 km/h
1.3 to 89.5 mph
1.2 to 77.8 knots
0 to 12 B |
0.6 to 60.0 m/s
118 to 11,811 ft/min
2.2 to 144.0 km/h
1.3 to 134.2 mph
1.2 to 116.6 knots
0 to 12 B |
1 inch|25 mm diameter impeller with precision axle
and low-friction Zytel® bearings. Startup speed stated as lower limit,
readings may be taken down to 0.4 m/s | 79 ft/min | 1.5 km/h | .9 mph | .8 kt
after impeller startup. Off-axis accuracy -1% @ 5º off-axis; -2% @ 10º; -3% @
15º. Calibration drift < 1% after 100 hours use at 16 MPH | 7 m/s.
Replacement impeller (NK PN-0801) field installs without tools (US Patent
5,783,753). Wind speed calibration and testing should be done with triangle
on impeller located at the top front face of the Kestrel. |
| Ambient Temperature |
|
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
0.9 °F
0.5 °C |
0.1 °F
0.1 °C |
-20.0 to 158.0 °F
-29.0 to 70.0 °C |
14.0.0 to 131.0 °F
-10.0 to 55.0 °C |
Hermetically-sealed, precision thermistor mounted
externally and thermally isolated (US Patent 5,939,645) for rapid response.
Airflow of 2.2 mph|1 m/s or greater provides fastest response and reduction
of insolation effect. Calibration drift negligible. Thermistor may also be
used to measure temperature of water or snow by submerging thermistor portion
into material -- remove impeller prior to taking submerged measurements and
ensure humidity sensor membrane is free of liquid water prior to taking humidity
based measurements after submersion. |
| Globe Temperature - Tg |
|
|
|
|
|
|
|
|
|
|
• |
|
|
°F
1.4 °C |
0.1 °F
0.1 °C |
-20.0 to 140.0 °F
-29.0 to 60.0 °C |
14.0 to 131.0 °F
-10.0 to 55.0 °C |
Temperature inside 1in|25 mm black powder coated
copper globe converted to Tg equivalent for standard 6 in|150 mm globe.
Closest equivalence obtained with airflow greater than 2.2 mph|1 m/s. |
| Relative Humidity |
|
|
|
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
3.0 %RH |
0.1 %RH |
5 to 95%
non-condensing |
0 to 100% |
Polymer capacitive humidity sensor mounted in
thin-walled chamber external to case for rapid, accurate response (US Patent
6,257,074). To achieve stated accuracy, unit must be permitted to equilibrate
to external temperature when exposed to large, rapid temperature changes and
be kept out of direct sunlight. Calibration drift +/- 2% over 24 months.
Humidity sensor may be recalibrated at factory or in field using Kestrel
Humidity Calibration Kit (NK PN-0802). |
| Pressure |
|
|
• |
|
• |
• |
• |
• |
• |
• |
• |
• |
• |
0.03 inHg
1.0 hPa|mbar
0.01 PSI
|
0.01 inHg
0.1 hPa|mbar
0.01 PSI |
8.86 to 32.49 inHg
300.0 to 1100.0 hPa|mbar
4.35 to 15.95 PSI
and
32.0 to 185.0 °F
0.0 to 85.0 °C |
0.30 to 48.87 inHg
10.0 to 1654.7 hPa|mbar
0.14 to 24.00 PSI
and
14.0 to 131.0 °F
-10.0 to 55.0 °C |
Monolithic silicon piezoresistive pressure sensor
with second-order temperature correction. Pressure sensor may be recalibrated
at factory or in field. Adjustable reference altitude allows display of
station pressure or barometric pressure corrected to MSL. Kestrel 4200
displays station pressure on a dedicated screen. Kestrel 2500 and 3500
display continuously updating three-hour barometric pressure trend indicator:
rising rapidly, rising, steady, falling, falling rapidly. Kestrel 4000
series displays pressure trend through graphing function. PSI display on
Kestrel 4000 series only. |
| Compass |
|
|
|
|
|
|
|
|
|
|
|
• |
• |
5° |
1°
1/16th Cardinal Scale |
0 to 360° |
0 to 360° |
2-axis solid-state magnetoresistive sensor mounted
perpendicular to unit plane. Accuracy of sensor dependent upon unit's
vertical position. Self-calibration routine eliminates magnetic error from
batteries or unit and must be run after every full power-down (battery
removal or change). Readout indicates direction to which the back of the unit
is pointed when held in a vertical orientation. Declination/variation
adjustable for True North readout. |
| CALCULATED
MEASUREMENTS |
| MEASUREMENT |
1000 |
2000 |
2500 |
3000 |
3500 |
3500
DT |
4000 |
4200 |
4250 |
4300 |
4400 |
4500 |
4500
HOR |
ACCURACY (+/-)* |
RESOLUTION |
SPECIFICATION RANGE |
SENSORS EMPLOYED |
NOTES |
| Air Density |
|
|
|
|
|
|
|
• |
• |
|
|
|
|
0.0002 lb/ft3
0.0033 kg/m3 |
0.001 lbs/ft3
0.001 kg/m3 |
Refer to Ranges for Sensors Employed |
Temperature
Relative Humidity
Pressure |
Mass of air per unit volume |
| Air Flow |
|
|
|
|
|
|
|
• |
|
|
|
|
|
6.71% |
1 cfm
1 m3/hr
1 m3/m
0.1m3/s
1 L/s |
Refer to Ranges for Sensors Employed |
Air Flow
User Input (Duct Shape & Size) |
Volume of air flowing through an opening.
Automatically calculated from Air Velocity measurement and user-specified
duct shape (circle or rectangle) and dimensions (units: in, ft, cm or m).
Maximum duct dimension input: 258.0 in | 21.5 ft | 655.3 cm | 6.55 m. |
| Altitude |
|
|
• |
|
• |
• |
• |
• |
• |
• |
• |
• |
• |
typical: 23.6 ft
7.2 m
max: 48.2 ft
14.7 m |
1 ft
1 m |
typical: 750 to 1100 mBar
max: 300 to 750 mBar |
Pressure
User Input (Reference Pressure) |
Height above Mean Sea Level ("MSL").
Temperature compensated pressure (barometric) altimeter requires accurate
reference barometric pressure to produce maximum absolute accuracy. Both
accuracy specs corresponds to a reference pressure anywhere from 850 to 1100
mBar. |
| Barometric Pressure |
|
|
• |
|
• |
• |
• |
• |
• |
• |
• |
• |
• |
0.07 inHg
2.4 hPa|mbar
0.03 PSI |
0.01 inHg
0.1 hPa|mbar
0.01 PSI |
Refer to Ranges for Sensors Employed |
Pressure
User Input (Reference Altitude) |
Air pressure that would be present in identical
conditions at MSL. Station pressure compensated for local elevation provided
by reference altitude. Requires accurate reference altitude to produce
maximum absolute accuracy. |
Crosswind &
Headwind/Tailwind |
|
|
|
|
|
|
|
|
|
|
|
• |
• |
7.1% |
1 mph
1 ft/min
0.1 km/h
0.1 m/s
0.1 knots |
Refer to Ranges for Sensors Employed |
Wind Speed
Compass |
Effective wind relative to a target or travel
direction. Auto-switching headwind/tailwind indication. |
| Delta T |
|
|
|
|
|
• |
|
|
|
|
|
|
|
3.2 °F
1.8 °C |
0.1 °F
0.1 °C |
Refer to Ranges for Sensors Employed |
Temperature
Relative Humidity
Pressure |
Difference between dry bulb temperature and wet bulb
temperature. When spraying, indicates evaporation rate and droplet lifetime.
Safe range for pesticide spraying is 4 to 16 °F / 2 to 9 °C. |
| Density Altitude |
|
|
|
|
|
|
• |
• |
• |
• |
• |
• |
• |
226 ft
69 m |
1 ft
1 m |
Refer to Ranges for Sensors Employed |
Temperature
Relative Humidity
Pressure |
Local air density converted to equivalent elevation
above sea level in a uniform layer consisting of the International Standard
Atmosphere. |
| Dewpoint |
|
|
|
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
3.4 °F
1.9 °C |
0.1 °F
0.1 °C |
15 to 95 % RH
Refer to Range for Temperature Sensor |
Temperature
Relative Humidity |
Temperature that a volume of air must be cooled to
at constant pressure for the water vapor present to condense into dew and
form on a solid surface. Can also be considered to be the water-to-air
saturation temperature. |
| Evaporation Rate |
|
|
|
|
|
|
|
|
|
• |
|
|
|
0.01 lb/ft2/hr
0.06 kg/m2/hr |
0.01 b/ft2/hr
0.01 kg/m2/hr |
Refer to Ranges for Sensors Employed |
Wind Speed
Temperature
Relative Humidity
Pressure
User Input (Concrete Temperature) |
The rate at which moisture is lost from the surface
of curing concrete. Requires user measurement and entry of concrete
temperature obtained with an accurate IR or probe thermometer (°F or °C, not
included). Readings should be taken 20 inches above pour surface with the
thermistor shaded, and averaged for 6-10 seconds using built-in averaging
function. |
| Heat Index |
|
|
• |
• |
• |
|
• |
• |
• |
• |
• |
• |
• |
7.1 °F
4.0 °C |
0.1 °F
0.1 °C |
Refer to Ranges for Sensors Employed |
Temperature
Relative Humidity |
Perceived temperature resulting from the combined
effect of temperature and relative humidity. Calculated based on NWS Heat
Index (HI) tables. Measurement range limited by extent of published tables. |
Moisture Content |
Humidity Ratio ("Grains") |
|
|
|
|
|
|
|
• |
• |
|
|
|
|
.3 gpp .04 g/kg |
0.1 gpp
0.01 g/kg |
Refer to Ranges for Sensors Employed |
Temperature
Relative Humidity
Pressure |
Mass of water vapor in a mass of air. |
| Relative Air Density |
|
|
|
|
|
|
|
|
• |
|
|
|
|
0.3% |
0.1% |
Refer to Ranges for Sensors Employed |
Temperature
Relative Humidity
Pressure |
The ratio, expressed as a percentage, of measured
air density to the air density of a standard atmosphere as defined by the
ICAO. |
| Thermal Work Limit (TWL) |
|
|
|
|
|
|
|
|
|
|
• |
|
|
10.9 W/m2 |
0.1 W/m2 |
Refer to Ranges for Sensors Employed |
Wind Speed
Temperature
Globe Temperature
Relative Humidity
Pressure |
Estimated safe maximum continuously sustainable
human metabolic rate (W/m2) for the conditions and clothing factors. Based
off of estimated metabolic output of typical human. On-screen zone warnings. |
Outdoor Wet Bulb Globe Temperature
(WBGT) |
|
|
|
|
|
|
|
|
|
|
• |
|
|
1.3 °F
0.7 °C |
0.1 °F
0.1 °C |
Refer to Ranges for Sensors Employed |
Wind Speed
Temperature
Globe Temperature
Relative Humidity
Pressure |
Measure of human heat stress defined as the
combination of effects due to radiation, convection, and conduction. Outdoor
WBGT is calculated from a weighted sum of natural wet bulb (Tnwb), globe
temperature (Tg), and dry bulb temperature (Td). User settable on-screen
warning zones. |
| Wet Bulb Temperature - Naturally Aspirated
(Tnwb) |
|
|
|
|
|
|
|
|
|
|
• |
|
|
1.4 °F
0.8 °C |
0.1 °F
0.1 °C |
Refer to Ranges for Sensors Employed |
Wind Speed
Temperature
Globe Temperature
Relative Humidity
Pressure |
Similar to psychrometric wet-bulb temperature (see
below). However, Tnwb only undergoes forced convection from the ambient air
velocity. Tnwb is a measure of the evaporative cooling that the air will
allow. This is accounted for by combining the effects of, mainly, relative
humidity and windspeed. |
| Wet Bulb Temperature - Psychrometric |
|
|
|
|
• |
• |
• |
• |
• |
• |
• |
• |
• |
3.2 °F
1.8 °C |
0.1 °F
0.1 °C |
Refer to Ranges for Sensors Employed |
Temperature
Relative Humidity
Pressure |
Temperature indicated by a sling psychrometer. Due
to nature of the psychrometric ratio for a water-air system, this
approximates the thermodynamic wet-bulb temperature. The thermodynamic
wet-bulb temperature is the temperature a parcel of air would have if cooled
adiabatically to saturation temperature via water evaporating into it. |
| Wind Chill |
|
• |
• |
• |
• |
|
• |
• |
• |
• |
• |
• |
• |
1.6 °F
0.9 °C |
0.1 °F
0.1 °C |
Refer to Ranges for Sensors Employed |
Wind Speed
Temperature |
Perceived temperature resulting from combined effect
of wind speed and temperature. Calculated based on the NWS Wind Chill
Temperature (WCT) Index, revised 2001, with wind speed adjusted by a factor
of 1.5 to yield equivalent results to wind speed measured at 10 m above
ground. Measurement range limited by extent of published tables. |
| ADDITIONAL
SPECIFICATIONS |
| Display
& Backlight |
• |
• |
|
• |
|
|
|
|
|
|
|
|
|
Reflective
3 1/2 digit LCD. Digit height 0.36 in / 9 mm. Aviation green
electroluminescent backlight. Manual activation with auto-off. |
| |
|
• |
|
• |
• |
|
|
|
|
|
|
|
Reflective
5 digit LCD. Digit height 0.36 in / 9 mm. Choice of aviation green or visible
red (NV models only) electroluminescent backlight. Manual activation with
auto-off. |
| |
|
|
|
|
|
• |
• |
• |
• |
• |
• |
• |
Multifunction,
multi-digit monochrome dot-matrix display. Choice of aviation green or
visible red (NV models only) electroluminescent backlight. Automatic or
manual activation. |
Response Time
& Display Update |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
All
measurements except those based on relative humidity respond accurately
within 1 second. Relative humidity and all measurements which include RH in
their calculation may require as long as 1 minute to fully equilibrate to a
large change in the measurement environment. Display updates every 1 second. |
| Max/Avg
Wind |
• |
• |
• |
• |
• |
• |
|
|
|
|
|
|
|
One-button
clear and restart of Max Wind Gust and Average Wind measurement. |
| |
|
|
|
|
|
• |
• |
• |
• |
• |
• |
• |
Max
and average wind calculation may be started and stopped independently of data
logging of other values, along with all other wind-related functions: air
velocity, crosswind, headwind/tailwind, wind chill, WBGT, TWL, evaporation
rate. |
| Data Storage & Graphical Display,
Min/Max/Avg History |
|
|
|
|
|
|
•
4000
points |
•
3200
points |
•
3200
points |
•
3600
points |
•
2300
points |
•
2900
points |
•
2500
points |
Minimum,
maximum, average and logged history stored and displayed for every measured
value. Large capacity data logger with graphical display. Manual and auto
data storage. Min/Max/Avg history may be reset independently. Auto-store
interval settable from 2 seconds to 12 hours, overwrite on or off. Logs even
when display off except for 2 and 5 second intervals (code version 4.18 and
later). Data capacity shown. |
| Data Upload & Bluetooth® Data Connect
Option |
|
|
|
|
|
|
• |
• |
• |
• |
• |
• |
• |
Requires
optional PC interface (USB or RS-232) or Bluetooth data transfer option and
provided software.
Bluetooth Data Transfer Option: Adjustable power consumption and radio range from up to 30 ft |
9 meters. Individual unit ID and 4-digit PIN code preprogrammed for easy
identification and data security when pairing and transmitting. Employs
Bluetooth Serial Port Protocol for data transmission. |
| Clock
/ Calendar |
• |
• |
• |
• |
• |
• |
|
|
|
|
|
|
|
Real-time
hours:minutes clock. |
| |
|
|
|
|
|
• |
• |
• |
• |
• |
• |
• |
Real-time
hours:minutes:seconds clock, calendar, automatic leap-year adjustment. |
| Auto
Shutdown |
• |
• |
• |
• |
• |
• |
|
|
|
|
|
|
|
After
45 minutes of no key presses. |
| |
|
|
|
|
|
• |
• |
• |
• |
• |
• |
• |
User-selectable
-- 15 or 60 minutes with no key presses or disabled. |
| Languages |
|
|
|
|
|
|
• |
• |
• |
• |
• |
• |
• |
English,
French, German, Italian, Spanish. |
| Certifications |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
CE
certified, RoHS and WEEE compliant. Individually tested to NIST-traceable
standards (written certificate of tests available at additional charge). |
| Origin |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
Designed
and manufactured in the USA from US and imported components. Complies with
Regional Value Content and Tariff Code Transformation requirements for NAFTA
Preference Criterion B. |
| Battery
Life |
• |
• |
• |
• |
• |
• |
|
|
|
|
|
|
|
CR2032,
one, included. Average life, 300 hours. Battery life reduced by backlight use
in 2000 to 3500 models. |
| |
|
|
|
|
|
• |
• |
• |
• |
• |
• |
• |
Standard Models: AAA Alkaline, two,
included. Average life, 400 hours of use, reduced by backlight or Bluetooth
radio transmission use.
Tactical Models: AAA Lithium, two, included. Average life, 400 hours of use,
reduced by backlight or Bluetooth radio transmission use. |
| Shock Resistance |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
MIL-STD-810g,
Transit Shock, Method 516.5 Procedure IV; unit only; impact may damage
replaceable impeller. |
| Sealing |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
Waterproof
(IP67 and NEMA-6). |
| Operational Temperature Limits |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
14°
F to 131° F | -10 °C to 55 °C Measurements may be taken beyond the limits of
the operational temperature range of the display and batteries by maintaining
the unit within the operational range and exposing it to the more extreme
environment for the minimum time necessary to take reading. |
| Storage Temperature |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
• |
-22.0
°F to 140.0 °F | -30.0 °C to 60.0 °C |
| Size
& Weight |
• |
• |
• |
• |
• |
• |
|
|
|
|
|
|
|
4.8
x 1.9 x 1.1 in / 12.2 x 4.8 x 2.8 cm, 3.6 oz / 102 g (including slip-on
cover). |
| |
|
|
|
|
|
• |
• |
• |
• |
|
• |
• |
5.0
x 1.8 x 1.1 in / 12.7 x 4.5 x 2.8 cm, 3.6 oz / 102 g. |
| |
|
|
|
|
|
|
|
|
|
• |
|
|
6.5
x 2.3 x 1.1 in / 16.5 x 5.9 x 2.8 cm, 4.4 oz / 125 g. |
| * NOTE: Accuracy calculated
as uncertainty of the measurement derived from statistical analysis
considering the comined effects from primary sensor specifications, circuit
conversions, and all other sources of error using a coverage factor of k=2,
or two standard deviations (2Σ). |
| Please note, these
specifications are valid for all Kestrel 4400 products and all other Kestrel
4000 series with a serial number higher than 659340. If your product has a
lower serial number, please reference the previous version
of the specifications. |
| © June 2012 Nielsen-Kellerman | 21 Creek
Circle Boothwyn PA 19061 USA | www.NKhome.com |
|
|
|
|
|
|
|
|
|
|